HL9900 Hall Effect Measurement

HL9900 is a turnkey, high performance hall effect system for the measurement of resistivity, carrier concentration, and mobility in semiconductors. Modular in concept, allowing easy upgrade paths, the system is suitable for a wide variety of materials, including silicon and compound semiconductors.

HL9900

HL9900 with original Nanometrics core technology inside

0.1 mΩ/sq to 100 GΩ/sq Sheet resistivity range, with HL9980
1 pA to 19.9 mA Source current range, with HL9980
90 to 500 K Temperature range with HL9950, ±0.5 K
0.5 T Permanent magnet field, 0.1% stability over 10 years

Models

One Hall System, Two Upgrade Modules

The HL9900 standard system covers low and high resistivity samples. The HL9980 buffer amplifier extends measurement to very high resistivity, and the HL9950 variable temperature unit adds measurement from 90 K to 500 K.

HL9900
HL9900 Hall effect measurement stage

Standard system

HL9900

Turnkey Hall system for sheet resistivities from 0.1 mΩ/square to in excess of 1 MΩ/square. Supplied with a two-temperature measurement stage complete with removable dewar and four micromanipulator probes.

Current range 100 nA to 19.9 mA
Input impedance 10¹⁰ Ω
Contact switching FET

Side by Side

Model HL9900 HL9980 HL9950
Role Standard system Buffer amplifier / current source Variable temperature unit
Current range 100 nA to 19.9 mA 1 pA to 10 µA —
Output impedance 10¹⁰ Ω 10¹³ Ω —
Input impedance 10¹⁰ Ω 10¹⁵ Ω ∥ 3 pF —
Input leakage 20 nA per input 70 fA per input —
Contact switching FET Dry reed —
Sheet resistivity 0.1 mΩ/sq to over 1 MΩ/sq Up to 100 GΩ/sq —
Temperature Two-temperature stage — 90 to 500 K

HL9980 and HL9950 are modules added to the HL9900 system. The HL9950 is fully compatible with the HL9980; with the buffer amplifier attached, the measurement circuit switches over automatically when sample current falls below 0.1 µA. On mobile, the table shows the model selected above.

Applications

What Hall Effect Measurement Determines

HL9900 has both low and high resistivity measurement capabilities.

  • Resistivity/Conductivity

  • Mobility

  • Bulk/Sheet carrier concentration

  • Doping type

  • Hall coefficient

  • Magnetoresistance

  • Vertical/Horizontal ratio of resistance

Features

Built for Fast, Accurate Hall Measurement

ASTM F-76 Sample Support

Van der Pauw, Hall Bar and Bridge measurements to ASTM F-76 standard, with correction for Hall scattering factor.

Simple Probe System

Simple probe system for convenient, fast sample throughput, in a compact bench top design. The measurement head is supplied with a two-temperature measurement stage complete with removable dewar and four micromanipulator probes.

Auto-Current Facility

A wide current range including auto-current facility in order to minimize sample heating.

Electric Field Limitation

User defined electric field limitation to avoid impact ionization effects at low temperatures.

AC/DC Measurement Modes

AC/DC measurement modes, using AC currents and phase sensitive detection.

Optional Buffer Amplifier

Optional high impedance buffer amplifier/current source to extend sheet resistivity measurements to 10¹¹ Ω/square.

Magnet

Permanent Magnet with Field Reversal

The HL9900/M50 uses a 0.5 T permanent magnet (nominal), with field reversal by magnet.

Type Permanent with field reversal by magnet
Field Strength ±1% of marked value
Pole gap 33.4 mm
Max. measurement diameter 25 mm
Stability 0.1% over 10 years

Software

No Programming Experience Required

With an intuitive user interface, no programming experience is required to set up or use the HL9900.

  • Van der Pauw, Hall Bar and Bridge measurements to ASTM F-76 standard

  • Correction for Hall scattering factor

  • Auto-current facility in order to minimize sample heating

  • User defined electric field limitation to avoid impact ionization effects at low temperatures

HL99WIN measurement window with resistivity, Hall and concentration results and a temperature scan table

HL99WIN results and temperature scan table

I-V curve check for contact pairs

I-V curve contact check

HL9950 cryo software with temperature control settings

HL9950 temperature control settings

Evaluate the HL9900 on Your Samples

Send samples to our Chicago lab. An applications engineer will measure them and recommend the right configuration.

Scroll to Top