Seebeck Effect Measurement System

The SB1000 Digital Seebeck Controller from Transient Signal Technologies delivers highly accurate, highly reproducible Seebeck coefficient measurements, using a differential technique with amplification inside the vacuum environment. Paired with the DC600 temperature controller, the system measures bulk materials, thin films, and nanomaterials from 80 K to 700 K.

TST Seebeck SB1000 system

SB1000 Digital Seebeck Controller by Transient Signal Technologies

70 nV Measurement resolution, gain 1000
80 to 700 K Temperature range with DC600
10 mW to 1 W Seebeck heater power range
2 to 8 mm Sample length, 2.5 mm wide or less

Temperature Options

From Room Temperature to 700 K

The SB1000 measures at room temperature on its own. Add a temperature controller with an LN2 cryostat, a hot stage, or both to measure across a wider range.

No temperature controller Room temperature only
DC600, LN2 cryostat 80 to 500 K
DC600, LN2 cryostat with hot stage 80 to 700 K
DC600 or K2000, hot stage system 300 to 700 K
Vacuum range 10 mTorr or better
Vacuum oil rotary pump 120 L/min or better

A Typical Variable Temperature System Includes

  • DC600 Programmable Temperature Controller and SB1000 Seebeck Effect Measurement Controller

  • Liquid nitrogen cryostats with integrated amplification circuitry into the vacuum environment

  • Kapton (80 K to 400 K) or ceramic (200 K to 700 K) sample mounting stages

  • Vacuum pump and vacuum accessories

  • Four soldering pins for electrical conductivity measurements by two probe or four probe technique

  • Dell Optiplex 7000 Tower desktop computer and 22” LCD monitor

  • All electronics mounted in 19” electronic rack

Applications

Metals to Semi-Insulators

Easy measurement of samples ranging from metals to semi-insulators, in single step operation against an internal reference material.

  • Bulk materials

  • Thin films

  • Nanomaterials

  • Thermoelectric materials research

Features

Built for High Signal-to-Noise Seebeck Measurement

Differential Measurement

Differential measurement technique of Seebeck coefficient with high signal-to-noise ratio.

In-Vacuum Amplification

Special amplification circuitry integrated into the vacuum environment allows high SNR.

Fast Data Acquisition

High data acquisition speed allows statistical measurements and analysis, with real-time thermal voltage transient measurements.

Variable Temperature Gradient

Vary the temperature gradient across the sample by applying selected power, 10 mW to 1 W, to the Seebeck heater.

Built-In Reference

Fully automated control with built-in reference, and single step operation against an internal reference material.

Compact Controller

Half-wide, full-length standard rack mount size, 5 lb (2.3 kg). Integrates into a rack mounted cabinet with a shelf.

Sample Stages

Low and High Temperature Mounting Stages

Each system is supplied with sample mounting stages for low temperature and high temperature operation. Each stage has an installed Seebeck heater for accurate temperature gradient control across the sample under test.

Kapton stage 80 to 400 K
Ceramic stage 200 to 700 K
Seebeck heater 10 mW to 1 W
Sample width 2.5 mm or less
Sample length 2 to 8 mm
Conductivity Four soldering pins for two or four probe measurement

Software

Automated Setup, Exportable Reports

A user-friendly, automated measurement setup with statistical data analysis. Reports are generated directly through the software suite.

  • Export to PNG, JPG, TXT, and CSV

  • Windows® 10 and Windows® 11

  • USB or serial port connection

  • 32-bit or 64-bit hardware

Software screenshot needed

Measurement setup

Software screenshot needed

Thermal voltage transient

Software screenshot needed

Statistical analysis and report